CKT1175-15 System

Stimulus / Measurement Units

Continuity Testing

continuity-testing

  • 12.5 mV – 50 VDC
  • 100µA – 2A
  • Resistance meas. as low as 1mΩ

Insulation Resistance Testing

insulation-resistence-testing

  • 12.5 mV – 2000 VDC
  • IR measurements to 5000 megohms
  • Hipot testing
  • Dwell time prog. from 1 ms to 1000s in 1 ms steps

AC Dielectric Testing

ac-dielectric-testing

  • 0.375-1500 VAC RMS
  • Test voltage ramp: 500V/0.1S, 500V/S, 500V/10S

AC/DC/I Measurements

ac-dc-i-measurement

  • voltage ranges – 100mV to 1000V
  • current ranges – 10µA to 100mA

Instrument Specifications


Continuity Test

TEST CURRENT

Full Scale Range

Programmable Steps

Accuracy

0.010 Amps

2.5µA

± 2%

0.100 Amps

25µA

± 2%

1.000 Amp

250µA

± 2%

2.000 Amps

2.5mA

± 2%

CONTINUITY TESTING COMPLIANCE VOLTAGE

Full Scale Range

Programmable Steps

Accuracy

12.5mVDC – 50.0VDC

12.5mVDC

± 2%

4-WIRE RESISTANCE RANGES

Current Range

Measurement Range

Accuracy

1mA – 10mA

100ohms – 50Kohms

± 2%

10mA – 100mA

10ohms – 5Kohms

± 0.5%

100mA – 1A

0.1ohm – 500ohms

± 0.5%, ± 1.5mohm**

1A – 5A

0.001ohm – 10ohms

± 0.1%,± 1mohm**

 

* *whichever is greater

 

2-WIRE RESISTANCE RANGES

Current Range

Measurement Range

Accuracy

1µA – 1mA

10Kohms – 500Kohms

± 2%

1mA – 10mA

1Kohms – 50Kohms

± 1.5%

10mA – 100mA

100ohm – 5Kohms

± 1.5%

100mA – 1A

10ohms – 500ohms

± 0.5%, ± 0.2ohm**

1A – 2A

1ohm – 50ohms

± 0.5%, ± 0.2ohm**

1A – 2A               (4-Pass Mode)

0.25ohms – 50ohms

± 0.5%, ± 0.05ohm**

0.1mA-1mA

>10ohms

±40%

 

** whichever is greater

DC Insulation Resistance Test

DC INSULATION RESISTANCE TEST VOLTAGE

Voltage Range

Programmable Steps

Accuracy

1VDC – 50VDC

12.5mVDC

± 2%

50VDC – 2000VDC

0.025VDC

± 2%

INSULATION RESISTANCE MEASUREMENT

Voltage Range

Measurement Range

Accuracy

1VDC – 50VDC

10ohms – 500Kohms

± 1%

50VDC – 500VDC

10Kohms – 500Megohms

± 1%

500VDC – 1500VDC

100Kohms – 1Gohm

± 1%

1500VDC-2000VDC

1Gohm-5Gohm

± 5%

AC Dielectric Breakdown Test

AC DIELECTRIC BREAKDOWN TEST VOLTAGE

Voltage Range

Programmable Steps

Accuracy

0.375VAC RMS -

1000VAC RMS

0.375 VAC RMS

± 2%

AC DIELECTRIC BREAKDOWN MEASUREMENT

Voltage Range

Measurement Range

Accuracy

100 VAC RMS -

1000VAC RMS

100µA-5mA

± 10%

Measurements

VOLTAGE MEASUREMENT

Full Scale Range

Accuracy

0.1V

± (0.5% ± 0.4 % FS)

1V

± (0.3% ± 0.3 % FS)

10V

± (0.25% ± 0.25% FS)

100V

± (0.25% ± 0.25% FS)

1000V

± (0.25% ± 0.25% FS)

CURRENT MEASUREMENT

Full Scale Range

Accuracy

10µA

± (0.4% ± 0.35 % FS)

100µA

± (0.3% ± 0.2 % FS)

1mA

± (0.3% ± 0.2% FS)

10mA

± (0.4% ± 0.2% FS)

100mA

± (0.4% ± 0.2% FS)

CAPACITANCE MEASUREMENT

Range

Accuracy

10pfd – 1nfd

± 3%, ± 5pfd **

1nfd – 100nfd

± 2.5%

100nfd – 5000µfd

± 2%

 

** whichever is greater